• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NORIS MEDICAL LTD TUFF TT IMPLANT; TUFF TT DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NORIS MEDICAL LTD TUFF TT IMPLANT; TUFF TT DENTAL IMPLANT Back to Search Results
Model Number NM-F4310
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 02/21/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The dhr was reviewed for lot#9000902 and there was no evidence discovered to indicate that a product defect or non-conformity contributed to the issue.To date, items of the same lot have been sold without any similar reported event during the period of review.The investigation for this event has not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Our manufacturing system assures, that production and process controls are in place to ensure that batches conform to the applicable specifications before they are distributed.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
A healthcare professional reported that nm-f4310 lot#9000902 implant lacked primary stability on tooth #10 due to bone too soft.A secondary implant was placed after removal.The implant was removed during implant placement with no report of observable clinical symptoms.According to the information, the patient is healthy.The device was forwarded to the manufacturer.No other medical issues were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUFF TT IMPLANT
Common Device Name
TUFF TT DENTAL IMPLANT
Manufacturer (Section D)
NORIS MEDICAL LTD
hataasia 8
nesher, 36888 08
IS  3688808
Manufacturer (Section G)
NORIS MEDICAL LTD
hataasia 8
nesher, 36888 08
IS   3688808
Manufacturer Contact
elena leonova
hataasia 8
nesher, 36888-08
IS   3688808
MDR Report Key21729427
Report Number3011390931-2025-00202
Device Sequence Number8734895
Product Code DZE
UDI-Device Identifier07290108812251
UDI-Public07290108812251
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K140440
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 03/31/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberNM-F4310
Device Catalogue NumberNM-F4310
Device Lot Number9000902
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2025
Initial Report FDA Received Date03/31/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age43 YR
Patient SexFemale
Patient Weight79 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-