| Catalog Number |
111645 |
| Medical Device Problem Codes |
Fracture (1260); Material Fragmentation (1261)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/06/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Tibial checkpoint broke off in patient when being removed.Half remains in patient.Additional information provided: part of the checkpoint remains within the patient¿s tibia.I am not aware of any adverse consequences.Surgeon has no plans to remove.
|
| |
|
Additional Manufacturer Narrative
|
|
As part of the normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted if additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
The event was not confirmed because the product was not available for inspection.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other complaints with similar event(s) for the lot referenced.Conclusions: the alleged failure mode was not confirmed because the product was not available for inspection.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
|
| |
|
Event or Problem Description
|
|
No new information.
|
| |
|