Visual inspection: during the investigation, no significant deformation or damage of the valve was determined.Permeability test: a permeability test has shown that the valve is permeable.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in the horizontal position.The results show that the valve operates within the accepted tolerances in the horizontal position.Adjustment test: the progav 2.0 was tested and is not adjustable.Braking force and brake function test: the brake functionality test has shown that the brake function is operational.However, the braking force cannot be measured due to the non-adjustability of the valve.Internal inspection: after dismantling of the valve, organic deposits were found in progav 2.0.Result: based on our investigation results, we can determine that the progav 2.0 is non-adjustable.The deposits visible in the valve have led to the functional impairment.Deposits of natural substances found in the body, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of organic deposits can impair the integrity of the valve.The examination of the return shipment is completed with the creation of this report.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|