|
The device was returned to olympus for inspection.A root cause could not be identified fore the burnt plastic distal end cover.Based on the results of the investigation, it is likely the following led to the malfunction: cause not established.The most probable cause of the failure related with white foreign materials is cause traced to failure to follow instructions.A labeling review was conducted using the product label pcf-ph190i.No anomalies were found.Is the reported risk/harm listed in the ifu? yes.Was the device used in accordance with the ifu/label? no, in the ifu it is stated: ¿for inspection of the objective lens cleaning function and inspection of the auxiliary water feeding function: nothing other than sterile water should be used for air/water feeding.No additives should be put into the sterile water.Non-sterile water may cause patient cross - contamination and/or infection.¿ is this a procedure problem where the uses of different materials than water could be sit in tubes and cause harm to the patient.Does the ifu/label need to be revised or was there an issue with translation, wording, or graphics? no.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|