| Model Number |
10FCC20 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/08/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product event summary: the 10fc20 sheath with lot number 0012273335 was returned and analyzed.Visual inspection of the handle area was performed.The inspection identified a kink at the side port tube.The steering mechanism and deflection test revealed the sheath didn¿t reach primary deflection at 90° and 180°.The dissection test revealed a broken pull wire inside the handle.In conclusion, the sheath failed the returned product inspection due to a side port tubing kink and a broken pull wire.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a pulsed field ablation procedure, when attempting to deflect into the left inferior pulmonary vein (lipv), an audible sound was heard that was surmised to be a pull wire snapping in the sheath.After that, the sheath was unable to deflect.The sheath was replaced which resolved the issue.The case was completed with pulsed field ablation.
|
| |
|
Search Alerts/Recalls
|