| Model Number |
M00545920 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0402 captures the reportable event of balloon burst.
|
| |
|
Event or Problem Description
|
|
Note: this report pertains to one of two devices that were used in the same patient and procedure.It was reported to boston scientific corporation that a hurricane rx dilatation balloon was attempted to be used in the common hepatic duct during a dilatation of the biliary tract procedure performed on (b)(6) 2025.During the procedure, the balloon burst at 6 atm when dilatation of the biliary stenosis was performed.The balloon was removed, and a second balloon was attempted to be used but also burst at 6 atm.It was reported that no pieces of the balloons detached inside the patient.The procedure was successfully completed with a third hurricane rx dilatation balloon.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|