W. L. GORE & ASSOCIATES, INC. GORE® DRYSEAL FLEX INTRODUCER SHEATH; INTRODUCER, CATHETER
|
Back to Search Results |
|
| Catalog Number |
DSF1433 |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Vascular Dissection (3160)
|
| Date of Event |
03/13/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The following information was reported to gore: on (b)(6) 2018, this patient underwent an endovascular treatment for a ruptured abdominal aortic aneurysm using gore® excluder® aaa endoprosthesis.On an unknown date, a proximal type i endoleak and a distal type i endoleak on the left side were identified.On (b)(6) 2025, a reintervention was performed using gore® dryseal flex introducer sheath and gore® excluder® aaa endoprosthesis.For the proximal type i endoleak, three aortic extender endoprosthesis were implanted proximally with the right femoral artery approach.For the left distal type i endoleak, using a 14fr sheath, with the left femoral artery approach, two contralateral leg endoprosthesis were implanted over the left external iliac artery after coil embolization of the left internal iliac artery.(this event is captured in: report# 3007284313-2025-03864 and 3013164176-2025-02454.) final angiography revealed dissection of the left external iliac artery near the distal end of the contralateral leg endoprosthesis.The dissection was decided to monitror and the procedure was completed.The physician reportedly stated as follows: as for the dissection, as the blood flow was not affected, i decided to monitor it.The patient's vessel was highly tortuous and that the dissection may have occurred during the stent grafts or the sheath manipulation.
|
| |
|
Additional Manufacturer Narrative
|
|
W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.H6: code d12: according to the gore® dryseal flex introducer sheath instructions for use (ifu), adverse events that may occur and / or require intervention include, but are not limited to: vascular trauma (i.E., dissection, rupture, perforation, tear, etc.).
|
| |
|