Patient presented in er from a gsw to distal humerus (b)(6) 2024.Taken to surgery for an orif with pangea extra-articular humeral plate and variax minifrag.Surgery was successful with post op x-rays to show correct alignment and utilization.Patient was discharged appropriately but then got into multiple altercations (fist fights) over the first few weeks' post op.Patient also stated he made attempts to remove his cast at home.He also tried to start a physical altercation with an ortho pa at his follow-up appointment.The patient was brought back to surgery for a revision on (b)(6) 2025.No infection was present.Hardware showed signs of failure with broken plate, broken screws, and a locking hole failure.Appears the locking screw migrated through the hole of the plate.
|
|
Please note the correction to g1 manufacturing site for devices and h6 clinical signs code.The reported event could be confirmed during the investigation.The device inspection revealed the following: the identification of the returned plate could be confirmed based on the catalog # and the lot # marked.The plate is broken in two parts at the 8th hole, both were returned for investigation.The breakage surface gives evidence of a sudden brittle fracture, resulting from excessive load.Furthermore, the plate is highly bent on both ends.Evidence indicates that this was done by hcp during implantation scratches visible on the implant's surface occurred most probably during the explantation.Dimensional inspection showed the device to be according to the specification.It must be noted that the inspection of the devices was only performed for procedural reasons.The information provided already gives the implant breakage."patient was discharged appropriately but then got into multiple altercations (fist fights) over the first few weeks post op.Patient also stated he made attempts to remove his cast at home.He also tried to start a physical altercation with an ortho pa at his follow-up appointment." the patient's post-operative behavior was undoubtedly the cause of the construction's failure, as it made any bone healing impossible.No orthopedic implant assembly could have been robust enough to withstand such stress in such a short time after implantation.As a reminder, violence should be avoided in all circumstances and at all times and the weeks following the implantation of an orthopedic device are no exception.The instructions for use clearly indicate the instructions for post-operative control and patient behavior required for a successful implantation and fracture healing.Fist fights and physical altercations with an orthopedic physician assistant are not among them.In addition, the plate was heavily bent by hcp.As a reminder, according to the instructions for use, bending, especially of this magnitude, should be avoided as much as possible, as it could reduce the fatigue strength of the device.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
|
|
Patient presented in er from a gsw to distal humerus (b)(6) 2024.Taken to surgery for an orif with pangea extra-articular humeral plate and variax minifrag.Surgery was successful with post op x-rays to show correct alignment and utilization.Patient was discharged appropriately but then got into multiple altercations (fist fights) over the first few weeks post op.Patient also stated he made attempts to remove his cast at home.He also tried to start a physical altercation with an ortho pa at his follow-up appointment.The patient was brought back to surgery for a revision on (b)(6) 2025.No infection was present.Hardware showed signs of failure with broken plate, broken screws, and a locking hole failure.Appears the locking screw migrated through the hole of the plate.
|