|
One used.List #a1141, was returned for evaluation.As returned no physical damage or anomalies were observed on the returned set, only tpn and lipids solution residual was observed.No mating device was returned for evaluation.The set was tested as per procedure and leak from the subassembly, clave® nano y-connector was observed.The sample was dissembled, and a damage/torn silicone seal condition was found.Complaint of leak can be confirmed.The probable of the puncture on seal is typical due to incompatible mating device during use, dfu states: the clave connector is compatible with luers with an internal diameter (id) between 0.062" and 0.110".The probable cause of the damage observed on the seal is unknown.A device history lot review of possible lot was performed and no anomalies, discrepancies and nonconformances were identified that might be related with the condition reported by customer.Updated information in sections: d1, d2, d4, g4, and h4.
|