• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. NIM® EMG - ENDOTRACHEAL TUBE; STIMULATOR, NERVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. NIM® EMG - ENDOTRACHEAL TUBE; STIMULATOR, NERVE Back to Search Results
Model Number 8229307
Medical Device Problem Code Unintended Deflation (4061)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/05/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the pre-inflation of the endotracheal tube before surgery, it was found that the cuff was leaking and could not be used normally.The problem was solved after replacing the endotracheal tube with a new one.There was no patient involved.
 
Additional Manufacturer Narrative
H3: analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3: product analysis found visually, there was no damage in the construction of the device observed by the unaided eye.Functionally, however, a syringe was used to inject 20cc of air into the cuff balloon and the cuff gradually deflated.For further analysis, a leak test was performed to localize the area of the leak and bubbles (leakage) were observed in distal portions of the cuff adjacent to the red electrode.Under magnification, there was a small puncture in the distal portions of the cuff.H6: fdm b21 and fdr c21 codes no longer apply.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H6: fdc d16 code no longer apply.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM® EMG - ENDOTRACHEAL TUBE
Common Device Name
STIMULATOR, NERVE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
justin ellis
6743 southpoint drive north
jacksonville, FL 32216
7635265677
MDR Report Key21736455
Report Number1045254-2025-01015
Device Sequence Number17139174
Product Code ETN
UDI-Device Identifier00763000882396
UDI-Public00763000882396
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K925640
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8229307
Device Catalogue Number8229307
Device Lot Number0228402997
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2025
Supplement Date Received by Manufacturer04/02/2025
06/13/2025
Initial Report FDA Received Date04/01/2025
Supplement Report FDA Received Date04/28/2025
06/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-