| Catalog Number |
03L80-22 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
|
| |
|
Event or Problem Description
|
|
The customer reported falsely decreased co2 results generated on the architect c4000 analyzer on a patient.The following results were provided: on (b)(6) 2025 = <10 meq/l (sample very lipemic), repeated sample on (b)(6) 2025 = <10 meq/l (triglyceride = 2728).On (b)(6) 2025, new sample = 13, another instrument (gem) = (b)(6).On (b)(6) 2025, new sample = 13, another instrument (gem) = 22, sent to labcorp = 23 (normal range: 22-27 meq/l).No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint investigation for falsely decreased co2 results included a review of data and information provided by the customer, labeling review, search for similar complaints, ticket trending review, and device history record review.Return testing was not completed as returns were not available.Trending review did not identify any trends for the issue for the product.Manufacturing documentation (device history record) for the likely cause did not identify any issues associated with the complaint issue.This indicates the assay is performing as expected.Labeling was reviewed and sufficiently addresses the customer's issue.Based on the available information, no deficiency was identified.
|
| |
|
Event or Problem Description
|
|
The customer reported falsely decreased co2 results generated on the architect c4000 analyzer on a patient.The following results were provided: on (b)(6) 2025 = <10 meq/l (sample very lipemic), repeated sample on (b)(6) 2025 = <10 meq/l.(triglyceride = 2728).On (b)(6) 2025, new sample = 13, another instrument (gem) = (b)(6).On (b)(6) 2025, new sample = 13, another instrument (gem) = 22, sent to (b)(6) = 23 (normal range: 22-27 meq/l).No impact to patient management was reported.
|
| |
|