| Catalog Number |
PLPUL2020 |
| Medical Device Problem Code |
Self-Activation or Keying (1557)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/18/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The reported device is being returned to conmed for evaluation.A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation.We will continue to monitor per regulatory requirements to help ensure patient safety.
|
| |
|
Event or Problem Description
|
|
The sales representative reported on behalf of the customer that the plpul2020, ultra surgical smoke evacuation pencil was being used on (b)(6) 2025 and there was ¿significant interference generated by the electrocautery, preventing the emg (evoked potential) device from functioning properly.Consequence: loss of confidence in the emg device, patient had to be awakened quickly to check for any impairments.Result: slight impairments, which may be temporary or permanent.Occurred at the end of a scoliosis case, during closure.Generator used ft10.Generator ft10 was activated for 5 minutes without noticing which interfered with ecg reading, account is wondering if there was a malfunction with electrosurgical pen as that activated itself or if the staff activated it without noticing because the energy alarm sound was too low on the generator.¿.The procedure was completed without the use of an alternate device.There was a delay of ¿30+ minutes¿.Further information has been requested; however, to date, a reply has not been received.This report is being raised due to the reported injury of slight impairments.
|
| |
|
Event or Problem Description
|
|
The sales representative reported on behalf of the customer that the plpul2020, ultra surgical smoke evacuation pencil was being used on (b)(6) 2025 and there was ¿significant interference generated by the electrocautery, preventing the emg (evoked potential) device from functioning properly.Consequence: loss of confidence in the emg device, patient had to be awakened quickly to check for any impairments.Result: slight impairments, which may be temporary or permanent.Occurred at the end of a scoliosis case, during closure.Generator used ft10.Generator ft10 was activated for 5 minutes without noticing which interfered with ecg reading, account is wondering if there was a malfunction with electrosurgical pen as that activated itself or if the staff activated it without noticing because the energy alarm sound was too low on the generator.¿.The procedure was completed without the use of an alternate device.There was a delay of ¿30+ minutes¿.Further information has been requested; however, to date, a reply has not been received.This report is being raised due to the reported injury of slight impairments.
|
| |
|
Additional Manufacturer Narrative
|
|
Examination of returned used device plpul2020 found that the device worked as intended.Examination was done with system 5000.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 15 reports, regarding 15 devices, for this device family and failure mode.During this same time frame (b)(4) have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, (b)(4).Per the instructions for use, the user is advised the following: inadvertent activation or movement of the activated electrode outside of the field of vision may result in injury to the patient.We will continue to monitor per regulatory requirements to help ensure patient safety.
|
| |
|
Search Alerts/Recalls
|