| Model Number |
NM-F3708 |
| Medical Device Problem Code |
Failure to Osseointegrate (1863)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/05/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A healthcare professional reported that nm-f3708 lot#9000778 implant lacked primary stability on tooth #7 due to bone too soft.A secondary implant was placed after removal.The implant was removed during implant placement with no report of observable clinical symptoms.According to the information, the patient is smoker, the patient has thyroid therapy.The device was forwarded to the manufacturer.No other medical issues were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The dhr was reviewed for lot#9000778 and there was no evidence discovered to indicate that a product defect or non-conformity contributed to the issue.The investigation for this event has not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Our manufacturing system assures, that production and process controls are in place to ensure that batches conform to the applicable specifications before they are distributed.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
| |
|
Search Alerts/Recalls
|