• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NORIS MEDICAL LTD TUFF TT IMPLANT; TUFF TT DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NORIS MEDICAL LTD TUFF TT IMPLANT; TUFF TT DENTAL IMPLANT Back to Search Results
Model Number NM-F5108
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/05/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The dhr was reviewed for lot#9001058 and there was no evidence discovered to indicate that a product defect or non-conformity contributed to the issue.The investigation for this event has not identified any signals indicating potential non-conformance's affecting the manufacturing and sterilization processes.Our manufacturing system assures, that production and process controls are in place to ensure that batches conform to the applicable specifications before they are distributed.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
A healthcare professional reported that nm-f5108 lot#9001058 implant lacked primary stability on tooth #14 due to bone too soft.A secondary implant was placed after removal.The implant was removed during implant placement with no report of observable clinical symptoms.According to the information, the patient is smoker, the patient has thyroid therapy.The device was forwarded to the manufacturer.No other medical issues were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUFF TT IMPLANT
Common Device Name
TUFF TT DENTAL IMPLANT
Manufacturer (Section D)
NORIS MEDICAL LTD
hataasia 8
nesher, 36888 08
IS  3688808
Manufacturer (Section G)
NORIS MEDICAL LTD
hataasia 8
nesher, 36888 08
IS   3688808
Manufacturer Contact
elena leonova
hataasia 8
nesher, 36888-08
IS   3688808
MDR Report Key21738539
Report Number3011390931-2025-00211
Device Sequence Number4022229
Product Code DZE
UDI-Device Identifier07290108812480
UDI-Public07290108812480
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K140440
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 04/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberNM-F5108
Device Catalogue NumberNM-F5108
Device Lot Number9001058
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2025
Initial Report FDA Received Date04/01/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
Patient SexFemale
Patient Weight48 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-