D10: (b)(6) - 300-30-06 - equinoxe preserve stem 6mm (b)(6) - 320-06-42 - glenosphere 42mm (b)(6) - 320-10-00 - equinoxe reverse tray adapter plate tray +0 (b)(6) - 320-15-01 - eq rev glenoid plate (b)(6) - 320-15-05 - eq rev locking screw (b)(6) - 320-20-00 - eq reverse torque defining screw kit (b)(6) - 320-20-30 - eq rev compress screw lck cap kit, 4.5 x 30mm (b)(6) - 320-20-30 - eq rev compress screw lck cap kit, 4.5 x 30mm (b)(6) - 320-20-38 - eq rev compress screw lck cap kit, 4.5 x 38mm.The product associated with the reported event is within the scope of recall z-1422-2024; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.The revision was most likely due to fracture of the compression screws from increased bending moments caused by baseplate loosening and/or an inadequate number of screws being used for fixation.Furthermore, the prothesis wear on the humeral liner from scapular notching and/or the inclusion of the polyethylene in the packaging recall could have contributed to the reason for revision.However, the cause for the wear and fracture can not be confirmed from the provided information.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|