MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK XS PT TEST 2X24 STRIPS; PROTHROMBIN TIME TEST STRIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK XS PT TEST 2X24 STRIPS; PROTHROMBIN TIME TEST STRIPS Back to Search Results
Catalog Number 04625315160
Medical Device Problem Code Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/05/2025
Type of Reportable Event Malfunction
Event or Problem Description
There was an allegation of questionable results from coaguchek xs meter, serial number (b)(6) compared to a patient's coaguchek xs meter with (b)(6) serial number.At 12:30 pm the meter result from the patient's coaguchek xs meter (b)(6) was 4.6 inr.The patient was using test strip lot 76900023 with expiration date 31-oct-2025.At 12:30 pm the meter result was 2.4 inr.The customer's therapeutic range is 2.0-3.0 inr.Customer tests on a weekly basis.
 
Additional Manufacturer Narrative
Coaguchek xs serial number is (b)(6).The product has not been received for further investigation at this time.If the product is returned in the future, a follow-up report will be submitted.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Section e3: occupation is patient/consumer.
 
Additional Manufacturer Narrative
The reporter¿s meter was provided for investigation where it was tested using retention strips and retention controls.Level 1 testing results (qc range = 1.0 - 1.4 inr): qc 1: 1.2 inr level 2 testing results (qc range = 2.4 - 3.6 inr): qc 2: 2.8 inr the obtained qc values were in the allowed range of the used combination strip lot - qc lot.All measurements were without error messages.The results alleged by the customer were not observed in the meter¿s patient result memory.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS PT TEST 2X24 STRIPS
Common Device Name
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key21739200
Report Number1823260-2025-00959
Device Sequence Number8734803
Product Code GJS
UDI-Device Identifier00365702128101
UDI-Public00365702128101
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K060978
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04625315160
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/05/2025
Supplement Date Received by Manufacturer04/16/2025
Initial Report FDA Received Date04/01/2025
Supplement Report FDA Received Date04/23/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age81 YR
Patient SexFemale
Patient Weight72 KG
-
-