| Catalog Number |
04625315160 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable results from coaguchek xs meter, serial number (b)(6) compared to a patient's coaguchek xs meter with (b)(6) serial number.At 12:30 pm the meter result from the patient's coaguchek xs meter (b)(6) was 4.6 inr.The patient was using test strip lot 76900023 with expiration date 31-oct-2025.At 12:30 pm the meter result was 2.4 inr.The customer's therapeutic range is 2.0-3.0 inr.Customer tests on a weekly basis.
|
| |
|
Additional Manufacturer Narrative
|
|
Coaguchek xs serial number is (b)(6).The product has not been received for further investigation at this time.If the product is returned in the future, a follow-up report will be submitted.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Section e3: occupation is patient/consumer.
|
| |
|
Additional Manufacturer Narrative
|
|
The reporter¿s meter was provided for investigation where it was tested using retention strips and retention controls.Level 1 testing results (qc range = 1.0 - 1.4 inr): qc 1: 1.2 inr level 2 testing results (qc range = 2.4 - 3.6 inr): qc 2: 2.8 inr the obtained qc values were in the allowed range of the used combination strip lot - qc lot.All measurements were without error messages.The results alleged by the customer were not observed in the meter¿s patient result memory.
|
| |
|