| Model Number |
AFAPRO28 |
| Medical Device Problem Codes |
Device Contamination with Body Fluid (2317); Gas/Air Leak (2946); Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product event summary: the data files were returned and analyzed.The reported issue was not observed in the log/data/multimedia files.No patient data file was received.The received failure data file did not contain any failure records for the date of the event.In conclusion, the reported issue (air ingress and blood in the coaxial umbilical cable) could not be assessed through data analysis.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a cardiac ablation procedure there was a pressure/flow issue.Air ingress was observed and there was blood found in the coaxial umbilical cable.The balloon catheter was replaced which resolved the issue.The case was completed with cryo.No patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|