| Catalog Number |
51-104170 |
| Medical Device Problem Code |
Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10:51-105170; item name tprlc xr t1 pps 17x154mm; lot # 3711480.51-105140; item name tprlc xr t1 pps 14x148mm; lot # 7017680.51-104160; item name tprlc 133 t1 pps ho 16x152mm; lot # 7216759.51-104170; item name tprlc 133 t1 pps ho 17x154mm; lot # 6002191.51-104110; item name tprlc 133 t1 pps ho 11x142mm; lot # 6858217.51-107150; item name tprlc 133 mp type1 pps ho 15.0; lot # 6081336.51-107170; item name tprlc 133 mp type1 pps ho 17.0; lot # 6663870.51-145150; item name tprlc xr mp t1 pps 15x115mm; lot # 6679519.51-107170; item name tprlc 133 mp type1 pps ho 17.0; lot # 7266249.51-145170; item name tprlc xr mp t1 pps 17x119mm; lot # 6862344.51-104150; item name tprlc 133 t1 pps ho 15x150mm; lot # 6669528.51-104110; item name tprlc 133 t1 pps ho 11x142mm; lot # 6283003.51-104100; item name tprlc 133 t1 pps ho 10x140mm; lot # 6908424.51-104170; item name tprlc 133 t1 pps ho 17x154mm; lot # 6572008.51-106160; item name tprlc 133 mp type1 pps so 16.0; lot # 6185837.G2: foreign: japan; customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that when inspecting the product, the sterile packaging was found damaged.There was no patient involvement.There is no further information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4;d9;g3;h2;h3;h6.Product was returned and evaluated.Visual evaluation of the returned product/provided photos identified damage to the sterile packaging (blister).Sterility had not been compromised.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Reported event is not related to a combination of products; therefore, a compatibility review is not applicable.The reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.The condition of the device when it left zimmer biomet is considered conforming to specification.The root cause of the reported event can be attributed to transit damage and a packaging design issue.The reported event has been confirmed by evaluation of the returned product and provided photos.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Search Alerts/Recalls
|