Catalog Number 200-02-38
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Pain (1994); Discomfort (2330); Swelling/ Edema (4577)
Date of Event 11/14/2023
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, approximately 9 years post op initial left tka, this 69 y/o male patient was revised due to a loose femur and recalled poly.Patient complained of pain.Patient was last known to be in stable condition following the event.Product not returning - chain of command.Due to recall hospital will not release.
 
Additional Manufacturer Narrative
D10 concomitants (b)(6) - 02-010-01-0250 - logic femoral ps cem left sz 5.(b)(6) - 02-012-45-5040 - lgc tibial fit tray cem sz 5f / 4t.H3: the reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.Operative notes and/or medical records were not provided for review of usage/ technique.A definitive root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The reason for the revision reported may be the result of the reported pain, femoral loosening, or due to inclusion of the polyethylene in the packaging recall.Prosthesis wear of the tibial insert could not be confirmed from the provided images.The minor wear of the patella was likely caused by high contact forces with the femoral component in vivo.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THREE PEG PATELLA 38MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville, FL 32653
3523771140
MDR Report Key21742034
Report Number1038671-2025-01706
Device Sequence Number18326794
Product Code JWH
UDI-Device Identifier10885862039613
UDI-Public10885862039613
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K932690
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/01/2019
Device Catalogue Number200-02-38
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2023
Supplement Date Received by Manufacturer07/17/2025
Initial Report FDA Received Date04/01/2025
Supplement Report FDA Received Date07/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/03/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization;
Patient Age69 YR
Patient SexMale
-
-