• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US MICROQA+ W/#4OC P3; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US MICROQA+ W/#4OC P3; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 212042
Medical Device Problem Code Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/19/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11: additional narrative: e1: the reporter¿s phone number was not provided.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Event or Problem Description
It was reported that during an unspecified surgical procedure, it was discovered that the microqa+ w/#4oc p3 device would not seat/advance.The reporter stated that during the procedure, an anchor was utilized, and all steps were followed according to the technique.However, when the suturing of the ligaments began, the anchor became dislodged from its body.As a result, another anchoring device had to be used, causing a delay of approximately fifteen minutes in the surgical procedure.The surgery was completed successfully without any impact on the patient.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11: additional narrative: investigation summary: the complaint device was not returned to j&j medtech orthopaedics, therefore unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device lot number (300l162), and no non-conformance was identified.Based on the current available information, we cannot determine a root cause for the reported failure.If the complaint device is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROQA+ W/#4OC P3
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key21745999
Report Number1221934-2025-01188
Device Sequence Number3809242
Product Code HWC
UDI-Device Identifier10886705001880
UDI-Public(01)10886705001880(10)300L162(17)270331
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K080352
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number212042
Device Lot Number300L162
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2025
Supplement Date Received by Manufacturer04/11/2025
Initial Report FDA Received Date04/01/2025
Supplement Report FDA Received Date04/21/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/30/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-