Catalog Number 320-15-01
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Aseptic loosening (4883)
Date of Event 03/15/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(d10) concomitant devices: 300-01-12 - equinoxe humeral stem primary press fit 12mm.320-42-00 - equinoxe reverse 42mm humeral liner +0.320-10-10 - equinoxe reverse tray adapter plate tray +10.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported that the patient underwent an initial total shoulder replacement on the right side, was revised for glenoid loosening approximately 6 years post-operatively, and underwent an arthroscopy and distal clavicle excision, approximately 11 months following the revision surgery.Subsequently, the patient has experienced aseptic glenoid loosening and developed severe pain associated with lifting activities.X-rays show evidence of glenoid component loosening.Component is clearly loose with a broken screw and radiolucency about the screws.There is possibility that may represent low grade infection.As a result, approximately 2.5 years after first revision, the patient underwent a second right shoulder revision.No further impact to the patient was reported.No other information is available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: d1/d2a/d2b, h6.Mdr section codes updated/corrected: b, c, d, e.The revision reported may be the result of compression screw loosening and fracture leading to glenoid loosening as reported.The reported radiolucency about the screws could indicate the screws were loose in the proximal region and still secured in the distal region.This would lead to a shear stress applied on the screws in the region still affixed in the bone during loading of the implant construct.This could lead to the reported screw fracture.However, the sequence of events and contributing factors to each cannot be determined from the information provided.Potential contributions from manufacturing, user, or patient-related contributing factors to the event cannot be evaluated as no radiographs, images, or clinical information was provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQ REV GLENOID PLATE
Common Device Name
PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
MDR Report Key21746128
Report Number1038671-2025-01713
Device Sequence Number3862664
Product Code KWT
UDI-Device Identifier10885862086488
UDI-Public10885862086488
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K063569
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number320-15-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2025
Supplement Date Received by Manufacturer06/30/2025
Initial Report FDA Received Date04/01/2025
Supplement Report FDA Received Date07/01/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization;
Patient Age67 YR
Patient SexMale
Patient Weight103 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-