Model Number 173046
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/11/2025
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, an endo paddle retractor was used during a laparoscopic adrenalectomy.During its use, the fabric piece came completely off of the paddle, inside of the patient.This was immediately noticed by operative staff, the fabric was removed from the patient in its entirety, and the device was removed from the sterile field.There was no patient injury.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6, h3.Evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted the paddle fully closed and opened by turning of black rotation knob.The device had difficulty retracting due to the damage to the paddle cover.It was reported that the fabric piece came completely off of the paddle, inside of the patient.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur when either the paddle retractor is retracted with the paddle in a partially open state; pushing the introducer forward over a paddle that is not fully closed may cause stretching and breakage of the introducer; or if the paddle retractor is subjected to excessive manipulation or used for leverage, these forces may cause the tearing of the cover/ cloth.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO PADDLE RETRACT
Common Device Name
RETRACTOR
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key21748613
Report Number2647580-2025-01023
Device Sequence Number13115251
Product Code GAD
UDI-Device Identifier10884523000771
UDI-Public10884523000771
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K914190
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 06/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number173046
Device Catalogue Number173046
Device Lot NumberP3F0066
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2025
Supplement Date Received by Manufacturer05/21/2025
Initial Report FDA Received Date04/01/2025
Supplement Report FDA Received Date06/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/05/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-