| Model Number |
MMT-377A |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Reference number (b)(4).Event occurred in the united states.It was reported that the patient faced insulin flow blocked alarm event on (b)(6) 2025.The blockage was in the tubing connector.No further information available.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: patient city - (b)(6).Patient country - united states.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information - this mdr is being submitted to include the below: h6: investigation results under type of investigation, investigation findings, investigation conclusions.H11: investigation summary: revision 21 of (b)(4) does not require a complaint that is type 2 reportable to open a child investigation.This child investigation was opened against a previous revision of (b)(4).Complaint investigation results: a complaint investigation has been initiated under complaint investigation child record (b)(4).The batch 6003364, in question was manufactured at the reynosa site.Threshold analysis: a query was run on 22-sep-2025 against "final reporting decision equal "serious injury" and "death", "lot number" criteria equal "6003364".The count of complaint is 0 which is below 3.No further statistical trending analysis is required.Device history record (dhr) review: the lot 6003364 was manufactured according to the work instruction (wi) version 15 and manufactured in the machine multivac 10 on 20/sep/2023, with a total of (b)(4) units.Glue-tubing lot: the lot 3j01361 was manufactured according to the (wi) version 10 and manufactured in the machine lc02 on 12/sep/2023, with a total of (b)(4) units.Review of the device history record (dhr) showed that all relevant tests required during the related processes had been fulfilled and met the requirements.No deviation was identified, nor maintenance events were recorded related to complaint code.Conclusion summary of complaint investigation: as a result of the following: no non-conformance (nc) raised during production related to complaint code, no further actions are required.This complaint will not require further root cause investigation nor corrective and preventive action (capa) plan.Therefore, this issue will be monitored through the post market surveillance activities.
|
| |
|
Event or Problem Description
|
|
To date no additional patient or event details have been received.
|
| |
|
Search Alerts/Recalls
|