WELCH ALLYN INC WIFI CSM,NIBP,MAS,RRP,BRAUNPRO6000,PLUG2; MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS)
|
Back to Search Results |
|
| Model Number |
75RE-2 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A picture of the plug was provided by the customer where there was visible burns marks on the pin.The connex spot monitors are intended to be used by clinicians and medically qualified personnel for monitoring of noninvasive blood pressure, pulse rate, noninvasive functional oxygen saturation of arteriolar hemoglobin (spo2), and body temperature in normal and axillary modes of neonatal, pediatric and adult patients.Monitoring respiration rate from photophlethysmogram (masimo rrp) is indicated for adult and pediatric patients greater than two years old.The most likely locations for patients to be monitored are general medical or surgical floors and general hospital and alternate care environments.This product is available for sale only upon the order of a physician or licensed health care professional.The 7000-ps is the connex spot monitor 35 watt power supply.The customer refused to return the unit for inspection, therefore a root cause into the reported malfunction could not be confirmed.Although there was no injury reported, if the pin on cable were to burn off, it could cause or contribute to a death or serious injury if this were to recur.Therefore baxter is reporting this device malfunction.
|
| |
|
Event or Problem Description
|
|
It was reported burn marks on cable pin.There was no reported injury.This report was filed in our complaint handling system as complaint # (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|