|
Block h6: imdrf device code a0501 captures the reportable investigation finding of electrode detached.Block h11: the returned orise proknife was analyzed, and a product analysis found the catheter was kinked.Functional inspection observed the electrode was broken and detached.The reported event of device failure to cut could not be confirmed.Based on all available information, it is likely that procedural factors, such as the length of time, power settings, handling technique, and tissue composition led to the damage of the electrode and subsequent use caused the electrode to detach.The kinks found on the catheter is likely due to the way in which the device was handled and manipulated during the procedure.Excessive manipulation of the device without enough care could have induced the defect found.Therefore, the most probable conclusion is adverse event related to procedure.Reported event of device failure to cut will be classified as cause not established due to this failure happened during the procedure and there is no objective evidence or descriptive conditions to establish the most probable cause.
|