• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INION OY INION FREEDOMPIN¿

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INION OY INION FREEDOMPIN¿ Back to Search Results
Model Number 1
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bacterial Infection (1735); Post Operative Wound Infection (2446)
Date of Event 02/10/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The situation was classified by customer as post-operative wound infection, however the cause of the event could not be determined.The possible causes were evaluated to potentially be related to the large trauma caused by the car accident and long operative time during the patient's original operation.Based on the information received by inion related to the event, it is possible that the patient's poor wound healing has been caused by late-onset wound infection approximately one year post-operatively.Late-onset wound infections are quite uncommon, but a recognized complication in patients after fracture fixation.Despite initial wound healing, the combination of significant soft tissue trauma, extended operative time, and implanted hardware may predispose to delayed surgical site infection.The presentation of infection after the wound has already seemed healed may reflect a previously subclinical or biofilm-associated infection that became clinically apparent over time.Alternatively, manifestation of infection due to any microorganism may be delayed because possible initial antimicrobial treatment did not result in complete microbial eradication.Bacterial infection is surgery-related risk, and this risk always exists when surgery is being performed, regardless which type of fixation implants are used.In case of infection, biofilm can form on both metallic and biodegradable implants.The inion implants are provided sterile and the batch record review which was carried out showed that the sterilization dose and dimensional and mechanical properties of the batch to which the implants used in the operations belonged to were within the specification.Blister and pouch seals of the inspected packages passed the inspection.Thus, the possible bacteria entering the body can be evaluated not to be related to inion implant sterility.Bioabsorbable implants can sometimes cause inflammatory tissue reactions related to the implant material degradation.Typically implant material degradation related inflammatory tissue reactions occur 1-2 years postoperatively.Time wise the patient's symptoms could also be related to implant material degradation, however it should be noted that implant material degradation related inflammatory tissue reactions are typically sterile fluid accumulations seen when the biodegradable implants start to degrade, there is no bacterial involvement in the degradation related inflammatory tissue reactions.But as the timing of the symptoms could also fit to material degradation related tissue reaction and since inion does not have bacterial culture results and cannot be fully aware of the course of the initial wound healing, given postoperative treatment and different aspects related to the emergence of wound infection type symptoms 1 year postoperatively after initial wound healing, a fully conclusive assessment whether the symptoms experienced 1 year after the implantation were caused by late onset bacterial infection or implant material degradation related inflammatory tissue reaction cannot be made with full certainty.However because the wound has now healed without the removal of the inion freedompin implant material (only metal implants were removed) and bacteria was found, at the moment the implant material degradation related tissue reaction in this case does not seem the most likely cause for the symptoms.Instead late onset bacterial infection seems more likely.Metal implants are well-known to harbor biofilm-forming bacteria, which can remain quiescent for long periods.These bacteria can be shielded from host immune responses and antibiotics, and may only become clinically apparent later.The resolution of symptoms following antibiotic therapy, debridement, and removal of only the metal implants supports the conclusion that metal implants were the nidus of the bacteria.Additionally, continued healing despite the retained bioabsorbable materials suggests that the bioabsorbable implants were not significantly contributing to the infection process.To conclude, the most probable cause of the late-onset wound infection type symptoms is a chronic, biofilm-associated infection of the metallic implants, manifesting after a period of dormancy.Clinical resolution following targeted treatment and selective hardware removal can be considered to support this assessment.However the implant material degradation related tissue reaction cannot be completely ruled out with the information received by inion, but as mentioned, in the implant material degradation related inflammatory tissue reactions there is no bacterial involvement.Also as the resolution of symptoms was achieved without the removal of bioabsorbable implants, it supports the conclusion that these materials did not contribute significantly to the poor wound healing symptoms either via material degradation or via harboring bacteria in the likely late-onset infection.
 
Event or Problem Description
A 56-year-old female with hypertension underwent open reduction and internal fixation of a tibia and fibula fracture.Both metal implants and absorbable inion freedompins (inion freedompin¿ kit ø2.0 mm and inion freedompin¿ kit ø1.5 mm (reported separately)) were used in the fixation.The patient was discharged successfully.Approximately 1 year later the patient was re-admitted to hospital with poor wound healing.Based on the information received, despite initial wound healing the wound started to have symptoms approximately 10 days prior to the re-admission to hospital (1 year after the original operation), bacterial cultures were taken and bacteria was found.The patient was treated with antibiotics, debridement and internal fixation removal of metal implants.The patient was discharged after wound healing.The inion freedompin implants were not removed and based on the information received by inion, the fractures have healed.
 
Additional Manufacturer Narrative
Item h6 (health effect - impact codes) was corrected, inion device was not removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INION FREEDOMPIN¿
Common Device Name
PIN
Manufacturer (Section D)
INION OY
lääkärinkatu 2
tampere, 33520
FI  33520
Manufacturer (Section G)
INION OY
lääkärinkatu 2
tampere, 33520
FI   33520
Manufacturer Contact
kati marttinen
lääkärinkatu 2
tampere, 33520
FI   33520
MDR Report Key21750901
Report Number3003407235-2025-00003
Device Sequence Number15512895
Product Code HTY
UDI-Device Identifier06438408004377
UDI-Public06438408004377
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K133932
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/01/2024
Device Model Number1
Device Catalogue NumberOPK-20K2
Device Lot Number2109060
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/02/2025
Supplement Report FDA Received Date04/23/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INION FREEDOMPIN¿ KIT Ø1.5MM (REPORTED SEPARATELY).; METAL IMPLANTS (UNKNOWN MANUFACTURER).
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient SexFemale
-
-