| Model Number |
MI1000 MED-EL CONCERT |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/31/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user's hearing performance is affected.Re-implantation is being considered.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance is affected.Re-implantation is being considered.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress appears very likely.However, to determine an exact root cause device investigation would be necessary.Re-implantation is considered but no date has been scheduled yet.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue wire breakages.The problems given in the recipient report appear to match the damage found.This is a final report.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance is affected.The user was re-implanted.
|
| |
|
Search Alerts/Recalls
|