• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OSCOR INC. SAFESHEATH® II; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OSCOR INC. SAFESHEATH® II; INTRODUCER, CATHETER Back to Search Results
Model Number SS6
Medical Device Problem Code Material Too Rigid or Stiff (1544)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/26/2025
Type of Reportable Event Malfunction
Event or Problem Description
Physician was performing an implant, and he noted the dilator was difficult to pass through the 6f safesheath but managed with a little more force than usually required.When breaking at the hub, he noted it took a little more effort than used to.The introducer was removed and a new safesheath from a different lot was used instead.After the procedure the physician cracked the hub of the discarded sheath out of interest and noted it was difficult to split.It appears the issue was with peeling the introducer.No additional procedure details were provided.No adverse patient effects were reported.No medical or surgical intervention was reported.There was no report that the procedure was not successfully completed.No pictures or videos related to the malfunctioned device are available.
 
Additional Manufacturer Narrative
No product was returned to oscor inc.For evaluation and the device was discarded; however, a complaint notification was provided.Without the return of the actual device in question for evaluation, oscor inc.Is unable to determine the exact cause for this incident.Based on the information provided by the supporting documentation, the physician noted the dilator was difficult to pass through the 6f safesheath but managed with a little more force than usually required.After the procedure the physician cracked the hub of the discarded sheath out of interest and noted it was difficult to split.A review of the manufacturing processes indicates the production operators are instructed to 100% visually inspect for any obvious defect prior to shipment.At this time, it is not possible to assign a definitive root cause for the event as reported.The complaint is non-verifiable as the product was not returned for evaluation.The instructions for use (ifu) in forms the user: flush sheath with 5cc of saline immediately before peeling sheath away to minimize back bleeding.Withdraw sheath and valve over the lead or catheter and from the vessel, while keeping the lead in place.Sharply snap the tabs of valve housing in a plane perpendicular to the long axis of the sheath to split the valve and peel sheath apart while withdrawing from the vessel.No action required, the complaint is non-verifiable as the product was not returned for evaluation.The event will be re-evaluated if additional information becomes available.Oscor will continue to monitor this event type and risk.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFESHEATH® II
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
OSCOR INC.
4875 palm harbor blvd.
palm harbor FL 34683
Manufacturer (Section G)
OSCOR INC.
4875 palm harbor blvd.
palm harbor FL 34683
Manufacturer Contact
ron greco
4875 palm harbor blvd.
palm harbor, FL 34683
MDR Report Key21751687
Report Number1035166-2025-00020
Device Sequence Number9700968
Product Code DYB
UDI-Device Identifier30891492002211
UDI-Public30891492002211
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K122084
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSS6
Device Catalogue NumberSS6
Device Lot NumberS9312090
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/05/2025
Initial Report FDA Received Date04/02/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/26/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-