(b)(4).Date sent: 4/2/2025 d4: batch # 547c65 investigation summary the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one pse60a device was returned with no apparent damage and with two ecr60g(b, c) reloads present.The reload (b) was received fully fired, the reload(c) was received partially fired 1/16, with the pan partially dislodged from the left proximal side, no driver was missing.The device was tested for functionality in the straight position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples meet the staple release criteria.It is possible that while loading the reload, the reload was pushed farther back than the reload alignment slot resulting in the knife pushing the one-piece sled forward and locking the reload.Please reference the instruction for use for more information.It is also possible that handling by the customer could dislodge the pan.Dislodgement as a result of the removal of the reload from the device is the most likely scenario.Please reference the instruction for use for more information.Device history review a manufacturing record evaluation was performed for the finished device batch number 547c65, and no non-conformances were identified.
|