|
Additional information: d2a, d2b, d9, g3, h2, h3, h6.One unpackaged ar-1317ft-1, batch 15296229, was received for investigation.Upon evaluation, the device's tip was damaged.Additionally, the suture assembly was returned torn/damaged.Functional testing could not be performed due to the damage to the device.The most likely cause of the reported failure is user error, such as misaligned insertion and/or prying/leveraging the device during use.Refer to investigation photos.The complaint allegation is confirmed.
|