C.R. BARD, INC. (BASD) -3006260740 POLYURETHANE, POWERPICC PROVENA SOLO2, 3CG; PERIPHERALLY INSERTED CENTRAL CATHETER (PICC)
|
Back to Search Results |
|
| Model Number |
N/A |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Each event reported to bd is evaluated and investigated in accordance with our complaint investigation procedures.The investigation process includes, but is not limited to, evaluation of the event details provided by the complaint facility, a review of complaint history, manufacturing records and risk document where applicable, and an evaluation of the subject device when available to identify potential contributing factors.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The device has not been returned to the manufacturer for evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer, both lumens would not flush or draw, both lumens treated with alteplase for 4 hours from 1100 the morning of this issue.The line would work if the skin was pulled toward the elbow and slightly up toward the bicep.Dressing change was done to try and reposition the external portion of the catheter, no change.The line was flushed as it was pulled, 1 cm at a time, requiring a new evl of 8cm to allow the line to function.Cxr was done and the tip is not at the subclavian vein/svc junction.Provider notified and replacement of the line was ordered.
|
| |
|
Search Alerts/Recalls
|
|
|