| Model Number |
N/A |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and it will be evaluated.A supplemental will be submitted with evaluation results.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer a piece broke off the end cap on huber needle line.All were caught before use on a patient.It was reported this occurred with five devices.This report addresses all five devices.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a damaged luer connector is confirmed and was determined to be supplier related.One 20g x 1.0in safestep infusion set was returned for evaluation.An initial visual observation revealed no obvious use residue in the sample.The luer adaptor was observed to be taped and a small white material was seen in its inner circumference.A microscopic observation revealed a chip on the inner stem of the dust cap.The chip appeared to have about the same size as the white material present in the inner circumference of the luer adaptor.The shape, location, and extent of the damage observed on the returned sample suggests the dust cap may have been damaged during the automated manufacturing process when the luer adaptor is closed with the dust cap.This complaint will be recorded for future trending and monitoring purposes.
|
| |
|
Search Alerts/Recalls
|