• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number SMPX685
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Contracture (1761); Deformity/ Disfigurement (2360)
Date of Event 02/17/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent a breast augmentation revision with mentor memorygel xtra breast implants 685cc and experienced capsular contracture baker grade iv on the left side post-operatively, which was confirmed during a physical exam.As a result, the patient is scheduled for removal on (b)(6) 2025.This report is for the right breast prosthesis.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.  as such, the investigation will be closed.  if the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.  this report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.  this report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.  if certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: anisomastia manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On may 15, 2025, additional information was received indicating the patient experienced bilateral capsular contracture.The baker grade was not identified for the right side.The replacement devices were identified as mentor gel breast implants (serial numbers (b)(6).On may 21, 2025, the mentor failure analysis lab received the device for evaluation.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.During visual evaluation, no apparent damage or visual anomalies were observed on the smooth mod + prof xtra 685cc returned device.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL XTRA BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key21754394
Report Number1645337-2025-03549
Device Sequence Number21257923
Product Code FTR
UDI-Device Identifier00081317025160
UDI-Public(01)00081317025160(17)260325(10)9554365
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/25/2026
Device Catalogue NumberSMPX685
Device Lot Number9554365
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/12/2025
Supplement Date Received by Manufacturer05/15/2025
Initial Report FDA Received Date04/02/2025
Supplement Report FDA Received Date06/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
Patient SexFemale
-
-