| Model Number |
G23163 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/21/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: device evaluation anticipated, but not yet begun.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
As reported, during an unknown procedure, the tip of a cxi support catheter separated.Additional information has been requested.
|
| |
|
Event or Problem Description
|
|
Additional information was received 01jul2025.As reported, during an angioplasty procedure in the leg, the "whole tip" of an 'cxi support catheter' separated as it was advanced over a competitor's wire.Reportedly, only the wire was in the patient but not the catheter.The catheter and the tip were removed from the wire and a new 'cxi support catheter' was opened to complete the procedure.A section of the device did not remain inside the patient's body.The patient did not require any additional procedures or experience any adverse effects due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, d9 (date returned to mfg), d10.Correction: h6 (annexes e and f).H3: device evaluated by mfg: device evaluation has begun; however, a conclusion is not yet available.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Summary of event: as reported, during an angioplasty procedure in the leg, the "whole tip" of a cxi support catheter separated as it was advanced over a competitor's wire.Reportedly, only the wire was in the patient but not the catheter.The catheter and the tip were removed from the wire and a new cxi support catheter was opened to complete the procedure.A section of the device did not remain inside the patient's body.The patient did not require any additional procedures or experience any adverse effects due to this event.Investigation evaluation: reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions, and quality control procedures were conducted during the investigation.A visual inspection of the complaint device was also conducted.The complaint device was returned to cook for investigation.The tip was not present on the catheter, and the separated tip was not returned.A document-based investigation evaluation was performed.A review of the device history record found two relevant non-conformances on eight devices from a sub-assembly lot; however, all affected product was scrapped prior to release from cook.A review of complaint history found no additional complaints for this lot number.The product ifu states ¿upon removal from package, inspect the product to ensure no damage has occurred.¿ a review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The information provided upon review of the dmr, dhr, ifu, and investigation of the returned device suggests that there is evidence the device was manufactured to specification.Although two relevant non-conformances were noted on a sub-assembly lot, all non-conforming product was scrapped prior to release from cook, there are 100% inspections in place to capture this non-conformance, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot-related complaints have been received from the field.Therefore, it was concluded that there is no evidence that additional non-conforming product exists in-house or in the field.Based on the information provided and the results of the investigation, cook has concluded that a component failure contributed to this event.It is possible that a wire guide could cause the tip of the catheter to separate; therefore, the tip could have been damaged when the catheter was loaded over the wire.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified, and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|