ALERE SAN DIEGO, INC. CARDINAL RAPID TEST HCG CASSETTE; VISUAL, PREGNANCY HCG, PRESCRIPTION USE Back to Search Results
Model Number FHC-102
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3 - date of event: was entered as (b)(6) 2025 as the customer was only able to indicate the events were around the week of (b)(6) 2025.Preliminary investigation findings: retained devices from the reported lot number were tested with hcg-negative clinical urine samples.The results were read at 3 and 4 minutes and all devices yielded valid negative results.No false positive results were observed during in-house testing.The case details were reviewed along with the complaint history for the reported issue and no indications of a systemic issue were identified.Manufacturing batch record review did not uncover any relevant non-conformances and found that the lot met quality control specifications.Review of the risk management report for this product found that the reported issue is within the risk profile for this device; no new hazard has been identified.A root cause could not be determined from the available information as the reported issue was not replicated during testing of retention product.Complaints are tracked and trended on a monthly basis.Per the package insert: - very low levels of hcg (less than 50 miu/ml) are present in urine specimen shortly after implantation.However, because a significant number of first trimester pregnancies terminate for natural reasons, a test result that is weakly positive should be confirmed by retesting with a first morning urine specimen collected 48 hours later.- a number of conditions other than pregnancy, including trophoblastic disease and certain non-trophoblastic neoplasms including testicular tumors, prostate cancer, breast cancer, and lung cancer, cause elevated levels of hcg.Therefore, the presence of hcg in urine specimen should not be used to diagnose pregnancy unless these conditions have been ruled out.- this test provides a presumptive diagnosis for pregnancy.A confirmed pregnancy diagnosis should only be made by a physician after all clinical and laboratory findings have been evaluated.Please note: urine specimens exhibiting visible precipitates should be centrifuged, filtered, or allowed to settle to obtain a clear specimen for testing.
 
Event or Problem Description
The customer reported receiving a total of 4 false positive hcg results using cardinal rapid hcg test cassettes for 4 patients.It was reported that these events took place sometime around the week of (b)(6) 2025, however specifics were unavailable.The third patient presented for a routine post-partum check-up.The patient was tested and a positive hcg result was obtained.A retest was performed using another device from the same kit which yielded a negative result.No adverse event reported.No further information was provided.See mdr 2027969-2025-00082 and 2027969-2025-00083 for patient 1 and 2 respectively.The positive result obtained by the fourth patient has been deemed not reportable as the customer deviated from the package insert, therefore a product malfunction was not determined, and no adverse events were reported.Therefore only 3 mdrs are being filed for this event.
 
Event or Problem Description
The customer reported receiving a total of 4 false positive hcg results using cardinal rapid hcg test cassettes for 4 patients.It was reported that these events took place sometime around the week of (b)(6) 2025, however specifics were unavailable.The third patient presented for a routine post-partum check-up.The patient was tested and a positive hcg result was obtained.A retest was performed using another device from the same kit which yielded a negative result.No adverse event reported.No further information was provided.See mdr 2027969-2025-00082 and 2027969-2025-00083 for patient 1 and 2 respectively.The positive result obtained by the fourth patient has been deemed not reportable as the customer deviated from the package insert, therefore a product malfunction was not determined, and no adverse events were reported.Therefore only 3 mdrs are being filed for this event.
 
Additional Manufacturer Narrative
B3 - date of event: was entered as 1mar2025 as the customer was only able to indicate the events were around the week of 9mar2025.H2, h3, h6 and h11 have been updated to include return testing.Investigation conclusion: retained and returned devices from the reported lot number were tested with hcg-negative clinical urine samples.The results were read at 3 and 4 minutes and all devices yielded valid negative results.No false positive results were observed during in-house testing.The case details were reviewed along with the complaint history for the reported issue and no indications of a systemic issue were identified.Manufacturing batch record review did not uncover any relevant non-conformances and found that the lot met quality control specifications.Review of the risk management report for this product found that the reported issue is within the risk profile for this device; no new hazard has been identified.A root cause could not be determined from the available information as the reported issue was not replicated during testing of retention and returned product.Complaints are tracked and trended on a monthly basis.Per the package insert: very low levels of hcg (less than 50 miu/ml) are present in urine specimen shortly after implantation.However, because a significant number of first trimester pregnancies terminate for natural reasons, a test result that is weakly positive should be confirmed by retesting with a first morning urine specimen collected 48 hours later.A number of conditions other than pregnancy, including trophoblastic disease and certain non-trophoblastic neoplasms including testicular tumors, prostate cancer, breast cancer, and lung cancer, cause elevated levels of hcg.Therefore, the presence of hcg in urine specimen should not be used to diagnose pregnancy unless these conditions have been ruled out.This test provides a presumptive diagnosis for pregnancy.A confirmed pregnancy diagnosis should only be made by a physician after all clinical and laboratory findings have been evaluated.Please note: urine specimens exhibiting visible precipitates should be centrifuged, filtered, or allowed to settle to obtain a clear specimen for testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDINAL RAPID TEST HCG CASSETTE
Common Device Name
VISUAL, PREGNANCY HCG, PRESCRIPTION USE
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
9942 mesa rim rd
san diego CA 92121
Manufacturer Contact
monika burrell
9942 mesa rim rd
san diego, CA 92121
8588052506
MDR Report Key21758851
Report Number2027969-2025-00084
Device Sequence Number18334590
Product Code JHI
UDI-Device Identifier20885380020251
UDI-Public(01)20885380020251(17)260327(10)0000857017
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K993317
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/27/2026
Device Model NumberFHC-102
Device Catalogue NumberB1077-22
Device Lot Number0000857017
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2025
Supplement Date Received by Manufacturer04/25/2025
Initial Report FDA Received Date04/02/2025
Supplement Report FDA Received Date04/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/28/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age34 YR
Patient SexFemale
Patient Weight89 KG
-
-