SMITH & NEPHEW, INC. Q-FIX KNTLS ASA 1.8MM W/1 MT SUT CB-BL; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
72205885 |
| Medical Device Problem Code |
Unstable (1667)
|
| Health Effect - Clinical Codes |
Perforation (2001); Device Embedded In Tissue or Plaque (3165)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unspecified procedure, after the 1.8 mm q-fix anchor was implanted, it did not hold tension as it should have.The repair suture was easily slick back and forth at will with little force on the repair suture.The anchor was left secured embedded into the patients bone.It was necessary to drill a backup bone hole to insert an anchor capable of maintaining tension.The procedure was completed with a s+n back up anchor, as it was put directly over the old anchor.There was a surgical delay of 10 minutes and no further complications were reported.
|
| |
|
Search Alerts/Recalls
|
|
|