The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the distal end with ccd module/us probe leak.Based on the result, we concluded that it was caused due to the physical damage applied on the distal end with ccd module/us probe.In addition, our technician confirmed that the us connector cable (long) buckled, the root brace rubber (insertion flexible tube) cut, and the objective lens scratched; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
|