| Model Number |
PED2-425-20 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received a report that the intracranial aneurysm embolization was performed.After the catheter was in place, the blood vessel diameter was measured and ped2-425-20 was selected for deployment.The stent tip could not be opened.After multiple attempts, it still could not be opened.The catheter was withdrawn and the operation was cancelled.No patient symptoms or further complications were reported as a result of this event.The pipeline was used for an approved (on-label) indication.The device and any accessories were prepared as indicated in the ifu.The patient was undergoing surgery for treatment of a saccular, unruptured ophthalmic artery aneurysm with a max diameter of 6 mm and a 5 mm neck diameter.The landing zone was 4.3 mm distal and 4.4 mm proximal.It was noted the patient's vessel tortuosity was minimal.Dual antiplatelet therapy (dapt) was administered.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information received reported the pipeline was placed in a straight segment when it failed to open.The pipeline was retrieved and deployed multiple times in an attempt to open the pipeline.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis ¿ as found condition: the pipeline flex shield was returned inside a biohazard bag, and a shipping box.¿ visual inspection/damage location details: the distal and proximal dps restraints were found to be intact.The dps sleeves were found intact with no signs of damage.The distal hypotube and ptfe shrink tubing were found to be intact with no signs of elongation.The braid's distal end appeared not opened and frayed.While the proximal end of the braid was found opened and frayed.No bends were found with the pushwire.No defects were found with the tip coil, distal marker, re-sheathing marker, re-sheathing pad, or the proximal bumper.No other anomalies were observed.¿ testing/analysis: n/a ¿ conclusion: based on the device analysis and reported information, the customer report of ¿failure/incomplete open distal¿ was confirmed as the braid's distal end was found not opened and frayed.The cause of the ¿failure/incomplete open¿ could not be determined.Possible causes of ¿failure/incomplete open¿ include vessel tortuosity, damaged braid, and deployment of the braid in the vessel bend.However, the customer indicated that vessel tortuosity was minimal, and the braid was not positioned in the vessel bend, which eliminates these factors out as potential causes.In this event, the damaged braid may have contributed to the failure to open.The braid can become damaged due to over-manipulation, deployment technique, delivering/retracting delivery wire against resistance, or deploying/resheathing the braid against resistance.However, the cause of the braid damage could not be determined.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|