| Model Number |
DVFB1D1 |
| Medical Device Problem Codes |
Failure to Deliver Shock/Stimulation (1133); Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
|
| Health Effect - Clinical Codes |
Ventricular Fibrillation (2130); Shock from Patient Lead(s) (3162)
|
| Date of Event |
03/17/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the implantable cardioverter defibrillator (icd) had inappropriately withheld therapy during episodes of ventricular fibrillation (vf) that were detected as a supra ventricular tachycardia (svt).In addition, there was possible right ventricular over-sensing following the shock, that did not appear to affect device function.The icd and right ventricular (rv) lead remains in use.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: b5, h6, additional codes.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the implantable cardioverter defibrillator (icd) had inappropriately delivered therapy during episodes of ventricular fibrillation (vf) that were detected as a supra ventricular tachycardia (svt).In addition, there was possible right ventricular over-sensing following the shock, that did not appear to affect device function. the icd and right ventricular (rv) lead remains in use. no further patient complications have been reported as a result of this event.
|
| |
|