H3: product analysis: an overheating test could not be conducted due to another device malfunction.The likely cause was identified as bearings worn.It was also noted noted drive shaft, input corroded.Color band faded.Not possible to insert/lock the bur.The tube bearings are corroded.Both flanged bearings and the non-flanged bearings are corroded.The input and output shafts are corroded.B3: date of notification: 2024-07-12 (date of event or estimated date of incident not provided to manufacturer).This regulatory report is being submitted due to a retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|