Model Number AA15
Medical Device Problem Codes Overheating of Device (1437); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Repair request initiated for device with the report of overheating and cosmetic deficiency.No patient impact reported.Repair request escalated to product event due to out of box failure.
 
Additional Manufacturer Narrative
H3: product analysis: evaluation could not be performed because of not possible to insert/lock the bur.The likely cause was identified as due to tube bearings and tube are corroded.It was also noted that both flanged bearings and the non-flanged bearings are corroded.The distal tube bearing is corroded and the other tube bearings are worn.B3: date of event or estimated date of incident not provided to manufacturer.This regulatory report is being submitted due to retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDAS REX LEGEND
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key21763546
Report Number1625507-2025-01222
Device Sequence Number8811775
Product Code HBB
UDI-Device Identifier00613994418852
UDI-Public00613994418852
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K020069
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAA15
Device Catalogue NumberAA15
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/12/2024
Initial Report FDA Received Date04/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/16/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-