Model Number UROLIFT 2 SYSTEM
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/29/2025
Type of Reportable Event Malfunction
Event or Problem Description
On january 29, 2025, neotract has been made aware that ¿during a prostatic urethral lift (pul) procedure, the first implant failed to deploy.The physician reported that the implant misfired, and the entire device had to be removed as the handle was not functioning.According to a neotract representative present during the procedure, there was a notable amount of bleeding as the physician extracted the device.The physician used graspers to remove the implant and opened a new handle to complete the procedure successfully.¿ additional information received on february 05, 2025, indicated that ¿according to the physician, the needle would not go back in, which is why the handle did not work.The patient's bleeding resolved over the weekend, and no intervention was needed.¿ on march 07, 2025, the investigation of the returned sample found that 3mm of the needle was not returned with the device.Further additional information received on march 19, 2025, indicated that ¿there was a small fragment of the needle that was removed from the patient.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEOTRACT UROLIFT 2 SYSTEM
Common Device Name
UROLIFT 2 SYSTEM
Manufacturer (Section D)
NEOTRACT, INC.
4155 hopyard road
pleasanton CA 94588
Manufacturer (Section G)
NEOTRACT, INC
4155 hopyard road
pleasanton CA 94588
Manufacturer Contact
laura gonzalez serrano
4155 hopyard road
pleasanton, CA 94588
9195448000
MDR Report Key21763619
Report Number3015181082-2025-00047
Device Sequence Number8591677
Product Code PEW
UDI-Device Identifier10814932020343
UDI-Public10814932020343
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K201837
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/17/2026
Device Model NumberUROLIFT 2 SYSTEM
Device Catalogue NumberIPN924184
Device Lot Number73D2400790
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2025
Initial Report FDA Received Date04/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/17/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-