Zimvie received one (1) bopt6510, (3i t3 tapered implant 6/5 x 10mm) for evaluation.Visual evaluation was performed, damage to the implant was identified.A third-party removal tool was stuck to the implant.Unable to remove it to inspect implant's drive feature.Dhr review was completed for the subject lot number 2120000481.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2120000481 for similar events and no other complaint was identified.Review completed utilizing keywords: ¿dental : functional : damaged drive feature¿.The customer did not submit images for the reported event.Based on the investigation and risk management file review as per rm-00053-haz, the most likely root cause determined from the investigation was customer error.Therefore, based on the available information, a device malfunction could not be verified.There implant had a third-party removal stuck inside the implant.Unable to confirm the damaged drive feature.The reported event was non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.The following sections have been updated: b4: date of this report.G3: date received by manufacturer.G6: checked "follow-up".H2: checked follow-up type.H3: changed "no" to "yes".H6: entered evaluation codes.H10: added manufacturer narrative.
|