| Catalog Number |
VCP752D |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 4/3/2025.H6 component code: g07002 - pending evaluation of returned device.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, the following was obtained: when did the needle detach or pull off from the suture (in the package, during removal from the package, during handling prior to using on the patient, or during use on the patient)? please provide details for each of the (b)(4) involved devices: during use.Not in the package.What is the procedure name and date? unk.Please provide the source or name and title of the external person providing answers to follow-up (external person submitting answers to sales rep): (b)(6).Please perform and document the follow-up attempt for product return.Please document the shipment tracking number in the notes or rmao sections: the device has been received at (b)(6) and will be shipped.Please check rmao.A device has been received; however, it has not yet been evaluated.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.During an unknown surgery, it felt that the needle-sutures were easier to detach than usual abnormally, so several replacements from the same lot were used in succession, but they were all detached easily.It was a control release needle.Another product was used to complete the case.There were no adverse consequences to the patient.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: h3, h6.Investigation summary: the product was returned to ethicon for evaluation.The returned sample shows that it was received one empty foil, a winding former labeled, and eight needle-suture combinations.The product code vcp752d is control release and require eight strands per packet.Upon a visual inspection of the returned sample, the swage and attachment area were noted to be as expected.Eight needle-suture combination was visually inspected, and no anomalies or issues related to pull-off were observed during the evaluation.The product code vcp752d contains an absorbable suture and the time of exposure of suture in the environment could not be determined; therefore, the functional test cannot be performed.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.No conclusion could be reached as to what caused the reported complaint.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|
| |
|
Search Alerts/Recalls
|
|
|