| Model Number |
LXMJ44L |
| Medical Device Problem Codes |
Peeled/Delaminated (1454); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a total prostatectomy with robot support, the device's tip insulation was detached when applied on periprostatic blood vessel.Another device was used appropriately with the same generator.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, g3.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a total prostatectomy with robot support, the device's tip insulation was detached and fell into p eriprostatic blood vessel.The broken piece was retrieved.Another device was used appropriately with the same generator.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d3 (mfr name and address), g3, h3, h6 h3.Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the jaw was damaged at the tip.The broken pieces were returned.Functional testing found that the damage to the device's seal plate likely occurred due to a drop or during the insertion or extraction of the device jaw from a trocar instrument.It was reported that the device's tip insulation was detached and fell into a periprostatic blood vessel.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: carefully insert and withdraw the instrument through the cannula to avoid damage to the device and or injury to the patient.Close jaws using device handle before insertion/extraction in the trocar.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|