• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. TRILOGY EVO; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. TRILOGY EVO; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number IT2110X21B
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/21/2025
Type of Reportable Event Malfunction
Event or Problem Description
A trilogy evo ventilator was returned to a third party service center for service.During the evaluation of the device at the third party service center a ventilator inoperative code was found related to the user interface software version string not matching the overall system version string.There was no reported patient involvement.The third party service technician observed the ventilator inoperative user interface software mismatch code in the device's event log.The technician replaced the device's system printed circuit assembly (pca) board to address the error code.In addition, a motor controller shutdown error code was also found in the device log.The blower was replaced to resolve the issue.A ventilator service required error code relating to an outlet pressure sensor railed to maximum negative value also appeared in the device's log.The replacement of the system printed circuit assembly (pca) board resolved the error.The air foam inlet filter, particulate filter, and muffler will be replaced due to the 4 year preventative maintenance procedure.No further investigation is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY EVO
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key21770364
Report Number2518422-2025-103685
Device Sequence Number9821984
Product Code CBK
UDI-Device Identifier00606959055421
UDI-Public(01)00606959055421
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryIT
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberIT2110X21B
Device Catalogue NumberIT2110X21B
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2025
Initial Report FDA Received Date04/03/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/19/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-