• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS INDIA LTD. VERADIUS; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS INDIA LTD. VERADIUS; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number VERADIUS UNITY
Medical Device Problem Code Radiation Output Failure (4027)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/05/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to philips that the wireless foot switch was not working.No harm was reported to philips.Philips has started an investigation of this complaint.
 
Additional Manufacturer Narrative
Philips has investigated this complaint.According to the additional information collected, the system was not in clinical use when the issue was identified.A philips field service engineer (fse) inspected onsite and verified that the foot switch stopped working since during fco implementation.Upon troubleshooting, fse replaced the wireless footswitch and base station, noting the footswitch battery was dead and the charger malfunctioned.To resolve the issue, fse replaced wireless footswitch and charger and return the part for further analysis, and it confirm that bracket was loose and for wfs base station no failure found.After replacement of wireless footswitch, base station and charger, the system was returned to use in good working order.At the time this complaint was received, philips did not have enough information to exclude the potential for death or serious injury on recurrence, and as such the complaint was reported.Since that time, it was identified that the issue occurred without patient involvement and was identifiable prior to procedure commencement, which has led to the determination that this is scenario is not likely to cause or contribute to death or serious injury if it were to recur.As per the instructions for use, the user of the product must institute a user routine checks program.The user of the product shall make sure that all checks and actions have been satisfactorily completed before using the product for its intended purpose.It is instructed to not use the product for any application until the user is sure that the user routine checks have been satisfactorily completed, therefore, as the device was outside of use at the time the issue was identified, the user would identify this issue prior to use and the issue happened due to improper maintenance.Based on re-evaluation, this case is not reportable.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERADIUS
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS INDIA LTD.
plot no. b-79, midc,
phase-ii, chakan
pune 41050 1
IN  410501
Manufacturer (Section G)
PHILIPS INDIA LTD.
plot no. b-79, midc,
phase-ii, chakan
pune 41050 1
IN   410501
Manufacturer Contact
marjan vos
plot no. b-79, midc,
phase-ii, chakan
pune 41050-1
IN   410501
MDR Report Key21771252
Report Number3003768277-2025-003096
Device Sequence Number15519784
Product Code OWB
UDI-Device Identifier00884838058194
UDI-Public(01)00884838058194
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K142708
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVERADIUS UNITY
Device Catalogue Number718132
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2025
Supplement Date Received by Manufacturer07/02/2025
Initial Report FDA Received Date04/04/2025
Supplement Report FDA Received Date07/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-