| Model Number |
LTF-190-10-3D |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/31/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that during reprocessing the image on the deflectable videoscope displayed a stripe noise.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Corrected fields: d9.H3.Additional information.The device was returned to olympus for inspection and the reported failure stripe noise image was confirmed.Based on the results of the investigation, and although we could not specify the root cause of the suggested event, it is likely the defect was caused by stress of repeated use, external factors, or handling.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|
Search Alerts/Recalls
|