| Model Number |
ARTHREX CHONDRAL DART, 18MM (5-BOX) |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an arthroscopic surgery after the device was implanted into the cartilage it fell out.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to use error.
|
| |
|