• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE REVISION MODULAR STOP; DRILL GUIDES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE REVISION MODULAR STOP; DRILL GUIDES Back to Search Results
Catalog Number 254600500
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/12/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.H3, h6: an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product or product id numbers are received at a later date, the investigation will be updated as applicable.
 
Event or Problem Description
It was reported that patient underwent total knee replacement on (b)(6) 2025.During the procedure, the modular stop sticks won't lock into place.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE REVISION MODULAR STOP
Common Device Name
DRILL GUIDES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key21773744
Report Number1818910-2025-05111
Device Sequence Number10161921
Product Code LXH
UDI-Device Identifier10603295478607
UDI-Public(01)10603295478607
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254600500
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/12/2025
Initial Report FDA Received Date04/04/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age74 YR
Patient SexMale
-
-